Agilent Technologies N9340A manual Using the Average Detector and Increased Sweep Time

Models: N9340A

1 172
Download 172 pages 20.6 Kb
Page 44
Image 44
Using the Average Detector and

3 Making Measurements

Using the Average Detector and

Increased Sweep Time

When the analyzer’s noise masks low- level signals, changing to the average detector and increasing the sweep time smoothes the noise and improves the signal’s visibility. Slower sweeps are required to average more noise variations.

1Refer to “Reducing Input Attenuation" on page 31, and follow steps 1, 2 and 3.

2Press [TRACE] > {More (1 of 2)} > {Detector} > {Average} to select the average detector.

3Press [BW/SWP] > {Sweep Time} to set the sweep time to 500 ms.

Note how the noise smoothes out, as there is more time to average the values for each of the displayed data points.

4Press [BW/SWP] > {Avg Type} to change the average type.

Figure 10 Using the average detector

34

N9340A User’s Guide

Page 44
Image 44
Agilent Technologies N9340A manual Using the Average Detector and Increased Sweep Time, Using the average detector