3 Making Measurements
Using the Average Detector and
Increased Sweep Time
When the analyzer’s noise masks low- level signals, changing to the average detector and increasing the sweep time smoothes the noise and improves the signal’s visibility. Slower sweeps are required to average more noise variations.
1Refer to “Reducing Input Attenuation" on page 31, and follow steps 1, 2 and 3.
2Press [TRACE] > {More (1 of 2)} > {Detector} > {Average} to select the average detector.
3Press [BW/SWP] > {Sweep Time} to set the sweep time to 500 ms.
Note how the noise smoothes out, as there is more time to average the values for each of the displayed data points.
4Press [BW/SWP] > {Avg Type} to change the average type.
Figure 10 Using the average detector