Remotely Mounting the Transmitter continued

Figure 3-7 Cable connections to sensors

ELITE and F-Series

sensor terminals

Flowmeter

cable

IFT9701

terminals

Before

Green

White

Brown

Violet

Yellow

Orange

Blue

Gray

Red

Brown

Red Clip drain wire back

Green

White Clip drain wire back

Blue

Gray Clip drain wire back

Orange

Violet

Yellow Clip drain wire back

Maximum cable length 1000 ft. (330 m)

Black

(Drains from all wire sets)

Brown

Red

Green 10

White 11

12

Blue GND

Gray

Orange

Violet

Yellow

You Begin

Prepare cable in accordance with the

instructions that are shipped with the cable

D and DL

Flowmeter

sensor terminals*

cable

 

Maximum cable length 1000 ft. (300 m)

Brown

Brown

Red

Red

Clip drain wire back

Orange

Green

White

Yellow

Clip drain wire back

Green

Blue

Blue

Gray

Clip drain wire back

Violet

Orange

Gray

Violet

Yellow

White

Clip drain wire back

Black

(Drains from all wire sets)

Brown

Red

Green

White

Blue

Gray

Orange

Violet

Yellow

IFT9701

terminals

10

11

12 GND

Getting Started

Prepare cable in accordance with the

instructions that are shipped with the cable

*Model D600 and DT sensors cannot be used with IFT9701 transmitters

Table 3-3 Sensor terminal designations

Terminal number

Wire color

Function

 

 

 

1

Brown

Drive +

 

 

 

2

Red

Drive –

 

 

 

3

Orange

Temperature –

 

 

 

4

Yellow

Temperature lead length compensator

 

 

 

5

Green

Left pickoff +

 

 

 

6

Blue

Right pickoff +

 

 

 

7

Violet

Temperature +

 

 

 

8

Gray

Right pickoff –

 

 

 

9

White

Left pickoff –

 

 

 

Mounting the Remote Transmitter

Power Supply and Output Wiring

Model IFT9701 Transmitter Instruction Manual

17

Page 25
Image 25
Emerson IFT9701 instruction manual Cable connections to sensors, Sensor terminal designations