Emerson IFT9701 instruction manual 1Location of LED, zero button, and LCD, 5.5Flowmeter zeroing

Models: IFT9701

1 90
Download 90 pages 24.24 Kb
Page 39
Image 39
Figure 5-1 Location of LED, zero button, and LCD

Flowmeter Startup continued

Figure 5-1 Location of LED, zero button, and LCD

Circuit board

Area approvals tag

compartment

User access not

 

normally required

 

Field wiring compartment

With optional LCD

Sensor wiring compartment

Zero button

Diagnostic LED

5.5Flowmeter zeroing

After the flowmeter has been fully installed, you must perform the zeroing procedure.

To perform the zeroing procedure using the flowmeter zero button, see the following instructions.

To use a HART Communicator for zeroing, see Chapter 6.

To use ProLink II software for zeroing, see Chapter 7.

5.5.1Zeroing procedure

1.Prepare the flowmeter for zeroing:

a.Install the flowmeter according to the instructions in this manual.

b.Apply power to the meter, then allow it to warm up for at least 30 minutes.

c.Run the process fluid to be measured through the flowmeter until the meter temperature approximates the normal process operating temperature.

d.Ensure that the sensor is completely filled with fluid.

2.Close the shutoff valve downstream from the meter.

5.5Flowmeter zeroing5.5.1Zeroing procedureManual background CAUTION

Flow through the flowmeter during flowmeter zeroing will result in an inaccurate zero setting.

Make sure fluid flow through the flowmeter is completely stopped during flowmeter zeroing.

Flowmeter Startup

Configuration with Communicator

Configuration with ProLink II

Troubleshooting

Model IFT9701 Transmitter Instruction Manual

31

Page 39
Image 39
Emerson IFT9701 instruction manual 1Location of LED, zero button, and LCD, 5.5Flowmeter zeroing, 5.5.1Zeroing procedure