Emerson IFT9701 4Configuration window - Frequency panel, 7.3Calibration procedures

Models: IFT9701

1 90
Download 90 pages 24.24 Kb
Page 59
Image 59
Figure 7-4 Configuration window – Frequency panel

Configuration with ProLink II Software continued

To configure the pulse output:

1.From the ProLink menu, click on Configuration.

2.Click on the Frequency tab. The Frequency panel appears, as shown in Figure 7-3.

Figure 7-4 Configuration window – Frequency panel

3.The pulse output produces a frequency proportional to the flow rate. Pulse output scaling requires entry of frequency and flow rate values. To scale the pulse output:

a.In the text box labeled Freq Factor, enter a value for the frequency in pulses per second (Hertz) that corresponds to the maximum flow rate.

b.In the text box labeled Rate Factor, enter the maximum value of the flow rate. The entered value corresponds to the frequency that was entered in the Freq Factor text box.

7.3Calibration procedures

Calibration procedures include zeroing and flow calibration.

7.3.1Auto zero

Flowmeter zeroing establishes flowmeter response to zero flow and sets a baseline for flow measurement.

Flowmeter Startup

Configuration with Communicator

Configuration with ProLink II

Troubleshooting

Model IFT9701 Transmitter Instruction Manual

51

Page 59
Image 59
Emerson IFT9701 instruction manual 4Configuration window - Frequency panel, 7.3Calibration procedures, 7.3.1Auto zero