Emerson IFT9701 instruction manual 6Totalizer Control window

Models: IFT9701

1 90
Download 90 pages 24.24 Kb
Page 62
Image 62
Figure 7-6 Totalizer Control window

Configuration with ProLink II Software continued

Use a mass flow unit during the calibration procedure. If the application requires volume flow measurement, choose a mass flow unit for the flow calibration, then choose a volume flow unit for the application.

To calculate the flow calibration factor:

1.Configure the transmitter to use a mass flow measurement unit (see Section 7.2.2).

2.In the Flow panel (depicted in Figure 7-2), enter the flowmeter’s 8-digit flow calibration factor into the Flow Cal text box. The flow calibration factor can be found on the flowmeter’s serial number tag, or on the certificate that is shipped with the sensor.

3.Zero the flowmeter (see Section 7.3.1).

4.With flow through the meter completely stopped, from the ProLink menu, click on Totalizer Control. The Totalizer Control window is displayed, as depicted in Figure 7-6.

Figure 7-6 Totalizer Control window

5. Click the Reset button to reset the internal totalizer to zero.

54

Model IFT9701 Transmitter Instruction Manual

Page 62
Image 62
Emerson IFT9701 instruction manual 6Totalizer Control window