20 to 30 VDC

6 watts typical, 14 watts maximum

Fused at 60 V/1.8 A Trip

Minimum startup voltage is 16 V at transmitter terminals

Maximum total resistance for wiring is 13 ohms

At startup, transmitter power source must provide a minimum of 0.7 amp of short-term current

A.3 Environmental limits

A.3.1 Temperature

Operating

See Process fluid vs. ambient temperature, below

Optional local display may become difficult to read below +14 °F (–10 °C).

Storage

–40 to +185 °F (–40 to +85 °C) without optional LCD

–4 to +58 °F (–20 to +70 °C) with optional LCD

A.3.2 Process fluid temperature limits

Integrally mounted transmitter

–40 to +257 °F (–40 to +125 °C)

See Process fluid vs. ambient temperature, below.

Remotely mounted transmitter

Refer to temperature limits for sensor.

Process fluid vs. ambient temperature

At their upper limits, process fluid temperature and ambient temperature restrict each other. The graphs provided below define the maximum recommended process fluid temperature based on maximum expected ambient temperature.