Troubleshooting continued

If the transmitter indicates an over range or sensor failure condition, follow these steps:

1.Isolate the transmitter from devices that use transmitter outputs to control the flow loop.

2.Refer to Table 8-5 to diagnose the problem.

3.If troubleshooting fails to reveal why over range and/or sensor failure messages have appeared, phone Micro Motion Customer Service. Phone numbers are listed on the title page of this manual.

Table 8-5 Troubleshooting over range and sensor failure conditions

Digital multimeter

 

 

(DMM)

Cause(s)

Corrective action(s)

• Open or short from red to

• Faulty cable

brown at transmitter

• Faulty drive coil in sensor

• Open or short from red to

• Moisture in sensor case or junction box

brown at sensor

 

 

 

• Open or short from green to

• Faulty cable

white at transmitter

• Faulty left pickoff in sensor

• Open or short from green to

• Moisture in sensor case or junction box

white at sensor

 

If open or short at transmitter terminals, check cable

If open or short at sensor junction box, return sensor to factory

Replace conduit and/or conduit seals

Repair cable

• Open or short from blue to

• Faulty cable

 

gray at transmitter

• Faulty right pickoff in sensor

 

• Open or short from blue to

• Moisture in sensor case or junction box

 

gray at sensor

 

 

 

 

 

• No open circuits

• Transmitter cannot calculate offset of flow

• Ensure that the sensor is filled with fluid

• No short circuits

signal

• Eliminate noise, then rezero

 

 

• Completely shut off flow, then rezero

 

 

 

 

• Transmitter cannot calculate flow rate

• Eliminate pipe stress, vibration, or

 

 

mechanical noise

Inappropriate density factor

Density < 0.200 g/cc or > 2.000 g/cc, volume flow is being measured

Density > 2.000 g/cc, mass flow is being measured

Erratic process density has caused flow tubes to stop vibrating

Plugged flow tube

Transmitter cannot calculate density

Characterize density values for sensor

Monitor density

Bring density within sensor limit

Purge flow tubes

Eliminate pipe stress, vibration, or mechanical noise

• Open or short from yellow

• Temperature outside sensor limit

to orange at transmitter

• Faulty cable

• Open or short from yellow

• Faulty lead length compensator

to orange at sensor

 

 

 

• Open or short from violet to

• Faulty cable

yellow at transmitter

• Faulty RTD in sensor

• Open or short from violet to

 

yellow at sensor

 

Bring temperature within sensor limit

Monitor temperature

If open or short at transmitter terminals, check cable

If open or short at sensor junction box, return sensor to factory

8.7Slug flow

Programmed slug flow limits minimize inaccurate measurement caused by significant quantities (slugs) of gas during liquid flow measurement or slugs of liquid during gas flow measurement. Such conditions adversely affect sensor performance by causing erratic vibration of the flow tubes, which in turn causes the transmitter to produce inaccurate flow signals.

62

Model IFT9701 Transmitter Instruction Manual

Page 70
Image 70
Emerson IFT9701 instruction manual Troubleshooting over range and sensor failure conditions, Slug flow