Emerson IFT9701 instruction manual 5Flow Calibration dialog box, Diagnosing zeroing failure

Models: IFT9701

1 90
Download 90 pages 24.24 Kb
Page 61
Image 61
Figure 7-5 Flow Calibration dialog box

Configuration with ProLink II Software continued

Figure 7-5 Flow Calibration dialog box

Diagnosing zeroing failure

If zeroing fails:

The Calibration Failure status light turns red.

The meter’s diagnostic LED blinks ON four times per second.

The flowmeter produces fault outputs.

The message “ELEC0” blinks on and off in the meter’s optional display. The most common sources of zeroing failure are:

Flow of fluid through meter during zeroing

Partially empty flow tubes

An improperly mounted meter

Re-zero the flowmeter after correcting the problem, or cycle power to the meter to abort the entire zeroing procedure and return to the previously established zero.

7.3.2Flow calibration procedure

Flow calibration involves adjusting the flow calibration factor so it accurately represents the sensitivity of the flowmeter to mass flow. Performing the flow calibration procedure in the field is optional.

Flow calibration is performed by running a batch of fluid through the meter, weighing the fluid, then using the Totalizer Control window to compare the weighed amount with the amount of fluid measured by the flowmeter. The accuracy of the scale used for weighing the fluid will determine the accuracy of the flow calibration. Use a scale that is highly accurate.

Flowmeter Startup

Configuration with Communicator

Configuration with ProLink II

Troubleshooting

Model IFT9701 Transmitter Instruction Manual

53

Page 61
Image 61
Emerson IFT9701 5Flow Calibration dialog box, Diagnosing zeroing failure, 7.3.2Flow calibration procedure