Pump maintenance

Battery

Certain features on the Paradigm pump use a lot of battery power.

Remote control

Backlight

Vibration alarm setting

If you use these features a lot, your battery will need to be replaced more frequently.

If your batteries have been in a cold place, (i.e., your car in the winter), allow the battery to warm to room temperature (up to 30 minutes) before inserting into the pump.

Installing a new pump battery

The Paradigm pump uses a AAA alkaline battery. As a safety mea- sure, Medtronic MiniMed has designed the pump to only accept a NEW battery. The pump is very particular, if you insert a used bat- tery, an alarm will be triggered.

CAUTION: It is highly recommended that you use an AAA alkaline Energizer battery. Do not use a carbon zinc battery in the pump.

Do not remove the battery unless you are changing your battery (installing a NEW battery). Replace it within one

(1)minute. If not replaced within one (1) minute, the screen may display an alarm message. Follow the instruc- tions in the message and make sure the time and date is set correctly. Install a new battery if the battery has been