4.The SET BASAL RATE screen will appear. Set the Basal Rates desired, then press ACT.

MiniMed

SET BASAL RATE 1

12:00A

A0. _

8 U

H

5.The BASAL RATE screen will appear. The screen will show the current rate, the time it started, and the 24-hour total.

The pump will return to the HOME screen if no buttons are pressed.

MiniMed

BASAL RATE A

Current Rate 0.8U/H Started #1 - 12:00A 24 Hr. Total 19.2U

6.The HOME screen will show open circles, indicating the pat- tern has been set. Check the status screen to verify that the pat- tern has been programmed.

This indicates

MiniMed

STATUS 9:12A

U100u

that Pattern A

Pattern A

 

is active.

Last Bolus: N 3.8U

 

Current basal

9:07am 11 NOV

 

Basal 1: 0.8 U/H

 

rate from

Units Remaining: 128U

Pattern A.

Battery: Normal

 

 

Remote: On

 

 

Tue 11NOV2000

 

 

S/N# 000000

 

72Chapter 5