Manual Prime
1.Make sure that the infusion set is disconnected from your pump to your body.
MiniMed
MANUAL PRIME
Disconnect from Body
Insert/Lock Reservoir
Hold ACT to Prime
2.Hold ACT to Prime. The pump will prepare to prime and will beep six times.
If you let go of the ACT button, either hold down ACT again to con- tinue or press ESC to repeat the entire “Rewind” procedure.
MiniMed
PREPARING TO PRIME
Hold ACT
3.Continue to hold the ACT button until you hear a second set of six beeps.
WARNING: DO NOT continue until you hear the second set of beeps and the screen from step 4 appears. Call the
28Chapter 2