Insulin delivery

Allows up to 48 individual basal rates to be programmed in a daily pattern.

Different basal rates can be programmed every 30 minutes.

Rates can be set from 0.0 to 35.0 units /hour in 0.1-unit increments (limited by maximum basal rate setting).

The basal delivery capacity is 0.0-35.0 units/hour.

Block

The block feature restricts access to all programming except suspend, selftest and the delivery of a bolus with the remote control.

The remote control must be used to program a bolus. The pump can be Suspended from either the remote or the pump.

Delivery accuracy

The pump has a delivery accuracy of +/- 5%.

Displacement accuracy

The pump has a displacement accuracy of +/- 2%.

Drive motor

The pump motor has a unique, patented design which delivers in precise increments.

Dual wave bolus

Delivers a normal bolus followed by a square wave bolus (limited by maximum bolus).