Normal bolus during a square wave bolus

Use this feature if you need additional insulin. You can program a normal bolus during the delivery of a square wave bolus. The nor- mal bolus will interrupt the square wave bolus that is being deliv- ered. The square wave delivery will continue once the normal bolus delivery is finished.

1.From the HOME screen, press ACT. Suspend will be selected. Select Bolus. Press ACT.

MiniMed

MAIN MENU

Bolus

Suspend

Basal

2. Set Bolus will be selected. Press ACT.

MiniMed

BOLUS MENU

Set Bolus

Bolus History

Max Bolus

3.The screen will now show that a square wave bolus is in progress. Press ACT.

MiniMed

 

SQUARE: Set

0.6U

Delivered

0.2U

Time Left

1:29

ACT to Program

 

44Chapter 4