Prime function

Is used to fill the infusion set tubing prior to use.

Primes are not counted in daily totals but are recorded separately in the prime history.

Fixed prime: 0.1-25.0 units (limited by maximum bolus).

Manual prime: unlimited. Fill rate is 1 to 5 units/second.

Program safety checks

Over 50 independent safety systems are continuously monitoring all pump operations.

Maximum infusion with single fault condition: 0.0074 ml.

Pump size

The Paradigm pump is 2 x 3 x .77 inches (.8 inches in battery area) [5 cm x 7.60 cm x 2 cm.]

Pump weight

The pump weighs less than 100 grams.

Remote control

Uses radio signals to allow users to program normal boluses or to suspend/resume their pumps.

Reservoir

Volume: up to 176 units of insulin.

Low reservoir alerts at approximately 30, 20, and 10 units of insulin remaining for U100. (20 and 10 units for U50)

The user-filled reservoir is made from impact-resistant, insulin-compatible polypropylene.

A clear window in the pump case allows users to see the remaining amount of insulin.