10.Press ESC to return to the INSULIN TYPE screen to change your Insulin Type if you have selected the wrong insulin type.

11.Press ACT to continue.

MiniMed

Consult your physician before changing insulin therapy.

12.Press ESC to return to the INSULIN TYPE screen or if you have consulted with your physician press ACT to verify the insulin change.

MiniMed

VERIFY INSULIN CHANGE

No U100u

Yes

13.Select Yes to verify the insulin change. Press ACT. Select No if you do not wish to change your insulin type.

14.The MANUAL PRIME screen will appear.

MiniMed

MANUAL PRIME

Disconnect from Body

Insert/Lock Reservoir

Hold ACT to Prime

WARNING: The pump must be rewound prior to inserting the new reservoir. Failure to disconnect and rewind the pump could result in an accidental infusion of insulin.

15. Be sure the infusion set is disconnected from your body.