Max delivery

This alarm alerts you when you have taken more insulin than expected based on maximum bolus and maximum basal rates.

Button error

This alarm will occur if a but- ton has been continually pressed for more than 3 min- utes.

MiniMed

3:25P

MAX

DELIVERY

Exceeded 1 hour max delivery Check BG ESC,ACT to clear

MiniMed

10:05A

BUTTON

ERROR

Button pressed for more than 3 min ESC,ACT to clear

Check settings

After the last button press, this alarm will repeat every 10 min- utes until the time and date are set. The following actions will cause this alarm to occur:

pump was reset due to an error,

battery has been taken out of the pump,

MiniMed

8.35A

CHECK

SETTINGS

Need to reprogram cleared settings or time without reprogramming ESC,ACT to clear

or the clear settings procedure performed.

You MUST reprogram the time/date and check to make sure all your settings; such as basal rate, are set as desired.

Alarms and Alerts 139