Delivering a square wave bolus

A square wave bolus delivers a bolus evenly over a period of time, from 30 minutes to 8 hours.

During a Square Wave Bolus other programming cannot be accom- plished, except for a Normal Bolus and Suspend.

1.From the HOME screen press ACT. Bolus will be selected. Press ACT.

MiniMed

MAIN MENU

Bolus

Suspend

Basal

2. Set Bolus will be selected. Press ACT.

MiniMed

BOLUS MENU

Set Bolus

Bolus History

Max Bolus

3. Select Square Wave Bolus. Press ACT.

MiniMed

BOLUS TYPE MENU

Normal Bolus

Square Wave Bolus

Dual Wave Bolus

42Chapter 4

Page 54
Image 54
Energizer MMT-511 manual Delivering a square wave bolus