MiniMed
These circles indicate Attention mode.

The following features or alarms turn on the open circles:

Dual/square wave bolus delivery

Basal patterns A or B

Temporary basal delivery

Block

Low reservoir

Low battery

Attention

Solid circles appear when insulin

delivery has stopped. This will hap- pen if operator interaction is

required, an alarm warning is received, or when the pump is sus-

pended. Alarm text; such as, Empty Reservoir will appear on the screen if activated. The warnings will be explained on the screen

when programming buttons are pressed. Use the to view all of

the instructions on the screen. Press ESC and then ACT to clear the alarm. Until the alarm is cleared, the pump is not delivering insulin.

Solid circles indicate that INSULIN DELIVERY HAS BEEN STOPPED.

The pump will beep/vibrate periodically when in the Attention or Special modes to remind you of the condition. The beep/vibrate frequency varies depending on the alarm.