It is recommended that you read your warranty statement included with your pump for information on what is covered during your warranty period.

Troubleshooting

My pump has a no delivery alarm

When a No Delivery alarm occurs, it means the pump is working correctly. Your pump is not broken, but it has detected that some- thing is preventing insulin from being delivered.

Do the following steps:

1.Check your blood glucose and take an injection if needed.

2.Make sure that there is insulin in your reservoir and the tub- ing is not kinked. If these are all right, go to step 5.

3.If necessary unkink tubing. Clear the alarm by pressing ESC and ACT. A screen will appear with two choices; Resume and Rewind. Select Resume.

4.If the reservoir is empty, clear the alarm by pressing ESC and ACT. Select Rewind and change your reservoir and infusion set per the instructions in Chapter 2, Getting Started.