4.The temporary basal rate will flash. Use the and but- tons to set the desired basal rate. Press ACT.

MiniMed

SET EDIT TEMP BASAL

03:30

0.6 UH-

Verifying temporary basal delivery

The pump will beep/vibrate three times every hour while deliver- ing temporary basal rate. You can also verify that a temporary basal rate is delivering by checking the STATUS screen.

1.From the HOME screen, press ESC to access the STATUS

screen. Use the and buttons to scroll through the information. Once you have completed your review, press ESC

to return to the HOME screen.

This shows

MiniMed

Temporary

STATUS 7:07A U100u

Basal is active

Temp Basal 0.6 U/H

at 0.6 U/H.

Duration 0:30

Time left 0:24

 

Last Bolus N 3.8U

Basal rate that

9:07A 11 NOV

would have

Basal 1: 0.8 U/H

Units Remaining: 128U

been delivered

Battery: Normal

if not for the

Tue 11NOV2000

Temporary

S/N# 000000

 

Basal rate.

 

In the example above the Temp Basal is programmed for 0.6 u/h for 30 minutes. The time left is 24 minutes, this indicates that 6 minutes have elapsed.

Basal 67

Page 79
Image 79
Energizer MMT-511 manual Verifying temporary basal delivery