4.The temporary basal rate will flash. Use the and
but- tons to set the desired basal rate. Press ACT.
MiniMed
SET EDIT TEMP BASAL
03:30
0.6 UH-
Verifying temporary basal delivery
The pump will beep/vibrate three times every hour while deliver- ing temporary basal rate. You can also verify that a temporary basal rate is delivering by checking the STATUS screen.
1.From the HOME screen, press ESC to access the STATUS
screen. Use the and
buttons to scroll through the information. Once you have completed your review, press ESC
to return to the HOME screen.
This shows | MiniMed | |
Temporary | ||
STATUS 7:07A U100u | ||
Basal is active | ||
Temp Basal 0.6 U/H | ||
at 0.6 U/H. | Duration 0:30 | |
Time left 0:24 | ||
| Last Bolus N 3.8U | |
Basal rate that | 9:07A 11 NOV | |
would have | Basal 1: 0.8 U/H | |
Units Remaining: 128U | ||
been delivered | Battery: Normal | |
if not for the | Tue 11NOV2000 | |
Temporary | S/N# 000000 | |
| ||
Basal rate. |
|
In the example above the Temp Basal is programmed for 0.6 u/h for 30 minutes. The time left is 24 minutes, this indicates that 6 minutes have elapsed.
Basal 67