Bolus stopped
If this alarm should occur, it is very important to check bolus history to review how much of the bolus was actually delivered. Reprogram a bolus with the amount not delivered, if required.
MiniMed
8.35A
BOLUS
STOPPED
Loose battery cap? Pump dropped or bumped?
Check bolus history, Reprogram bolus
if required ESC,ACT to clear
Failed batt test
The pump is designed to test each new battery installed in the pump. This is a safety fea- ture to make sure that a battery that already has low voltage is not used. If the battery does not pass the test this alarm will occur.
8.35A
FAILED
BATT TEST
Replace with new battery Delivery stopped ESC,ACT to clear
Always make sure that you
install a NEW battery into the pump.
This alarm will also happen if a used battery is installed in the pump or if the installed battery is removed and then put back in.