Energizer MMT-511 manual Maximum bolus limit max bolus, When square/dual wave bolus option is on

Models: MMT-511

1 170
Download 170 pages 14.39 Kb
Page 62
Image 62

When square/dual wave bolus option is on

1.From the HOME screen, press the button.

2.The Bolus Type menu will appear.

MiniMed

BOLUS TYPE MENU

Normal Bolus

Square Wave Bolus

Dual Wave Bolus

3.Select the desired bolus type. Press ACT. The screen that appears will be different, depending on the type of bolus you selected. Program your desired settings.

Maximum bolus limit (max bolus)

A maximum bolus limit sets a safety limit for the amount of insulin that can be delivered in a single bolus.

It’s important to discuss this feature with your healthcare profes- sional to decide on the maximum bolus amount of insulin that can be delivered at one time. The Maximum Bolus Limit can be set from 0.0 to 25.0 units. The factory setting is 10.0 units.

Setting the maximum bolus limit

1.From the HOME screen, press ACT. Bolus will be selected. Press ACT.

MiniMed

MAIN MENU

Bolus

Suspend

Basal

50Chapter 4

Page 62
Image 62
Energizer MMT-511 Maximum bolus limit max bolus, When square/dual wave bolus option is on, Setting the maximum bolus limit