15 Ethernet interface

- Decimal digital data:

Made of a succession of ASCII-coded characters beginning with a digit or a sign (+ or -) It can be of NR1 (integer), NR2 (decimal) or NR3 type (with an exponent).

For example for a digital character: -25.02.

- Text data:

7-bit ASCII-coded character chain between quotation marks (") or apostrophes ('). For example: "A".

-Any succession ASCII characters: ends with the termination sequence.

14.3.Standard instructions

All these instructions begin with "*".

*IDN ? REQUEST FOR IDENTIFICATION OF AN APPLIANCE answer by the appliance: 4 pieces of data separated by ',':

-the trademark of the appliance

-the name of the appliance, followed by _nn where nn is the number of inputs of the recorder

-the serial number of the appliance (0 if unknown)

-the version number of the software: x.xx x

*OPT ? REQUEST FOR IDENTIFICATION OF THE OPTIONS OF AN APPLIANCE answer by the appliance: n pieces of data separated by ',':

-number of input modules

-number of channels per module

*RST RESET OF AN APPLIANCE

action: initialisation of the recorder in a fixed configuration (voltage inputs, 10V range, 0V centre...)

*REM SWITCHING TO PROGRAMMATION (REMOTE) required in RS232C before sending any other programming command.

*LOC SWITCHING TO LOCAL MODE

*CLS DELETING THE STATUS REGISTERS action: the appliance resets the status registers to zero.

*ESE VALIDATION OF THE STANDART EVENT BITS OF AN APPLIANCE *ESE is followed with a digit, from 0 up to 255

action: changes the validation register of the standard events and updates the ESB bit in the status register of the service request (see the following paragraph).

*ESE ? INTERROGATION OF THE CONTENT OF THE VALIDATION REGISTER OF THE STANDARD EVENTS OF AN APPLIANCE

answer by the appliance: NR1 number between 0 and 255 (see the following paragraph).