16 Technical specifications

16.11.4. Security isolation class

installation category

 

 

 

 

 

 

 

 

 

 

 

 

Designation

Accordance test

 

Specifications

Limits

 

Criteria

 

 

 

 

30 MHz to 230 MHz

Measure distance 3 meters

 

 

Radiated Emissivity rayonn e

NF EN 55022

 

40 dB V/m

 

 

 

 

230 MHz to 1 GHz

 

Class B

 

 

 

 

47 dB V/m

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0,15 MHz to 0,5 MHz

66 to 56 dB V QP

 

 

Conducted Emissivity

 

 

56 dB

V QP

 

 

 

NF EN 55022

 

0,5 MHz to 5 MHz

 

Class B

 

(lines, power supply)

 

60 dB

V QP

 

 

 

 

5 MHz to 30 MHz

 

 

 

 

 

 

(mean value = QP - 10dB)

 

 

 

 

 

 

 

 

Harmonic Distortion

CEI 61000-3-2

 

 

Picture 1 of

the standard

Class A

 

 

 

 

0 to 2 kHz

 

 

 

 

 

Variation of tensions and flickers

CEI 61000-3-3

 

chap 5 of the standard

-

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Air

 

 

 

 

 

 

CEI 61000-4-2

 

Contact

N = +/- 4 kVolts

 

Criteria B

 

 

 

 

discharges

N = +/- 8 kVolts

 

 

 

 

 

 

 

 

 

 

 

 

Enveloppe immunity

 

 

80MHz to 1 GHz

10 V/m without mod.

 

 

CEI 61000-4-3

 

1,4 GHz to 2 GHz

Criteria A

 

 

 

 

AM 80% 1000Hz

 

 

 

 

 

 

CEI 61000-4-8

 

50 Hz

30 A/m

 

Criteria B

 

 

 

 

 

 

 

 

 

 

 

 

Line energy +/- 2 kV

 

 

 

CEI 61000-4-4

 

5-50 ns / 5 kHz

Line Ethernet +/- 1 kV

Criteria B

 

 

 

Line measure input +/- 1 kV

 

 

 

 

 

 

 

 

 

 

 

Line earth +/- 1 kV

 

 

 

 

CEI 61000-4-5

 

1,2 / 50 s (8/20)

line / line +/- 1 kV

 

Criteria B

 

 

 

line / earth +/- 2 kV

 

Access immunity

 

 

 

 

 

 

 

 

Line energy

 

3V

 

 

 

 

 

 

 

 

 

 

CEI 61000-4-6

 

150 kHz to 80 MHz

Line Ethernet

without

Criteria A

 

 

 

AM 80% 1000 Hz

Line input measure

mod.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CEI 61000-4-11

 

Reduction 100%

0,5 cycle with each polarity

Criteria B

 

 

 

 

 

 

 

 

 

Class 1 product

 

 

 

 

 

 

 

Security

conform with EN61010-1

 

 

 

 

Pollution degree

2

 

 

 

 

 

 

Installation category (overvoltage category)

 

 

 

 

 

Mains input

category II

 

 

 

 

 

 

 

Measurement input

category III 600 V, overvoltage 6000 V

 

 

 

16.11.5. Electromagnetic compatibility

Assessment criteria for the functioning that were applied for the test: Criteria A : Normal behaviour within the limits of the specifications Criteria B : Temporary decrease or self-recovered loss of function

Criteria C : Temporary decrease or loss of function or behaviour that requires the intervention of the user or a reset of the system

Particular precautions are necessary to preserve the conformity of the product, particularly the use of shielded cables.

Page 8

Page 127
Image 127
Escort 3008B, 3016B, 3004B user manual Electromagnetic compatibility, Accordance test