16 Technical specifications

16.1.4.Temperature recording

CAPTEUR

using environment

Ranges

 

 

 

PT100

-200 C to 850 C

20 C to 1000 C

COUPLE J

-210 C to 1200

C

20 C to 2000 C

CO UPLE K

-250

C to 1370

C

20 C to 2000 C

COUPLE T

-200 C to 400 C

20 C to 500 C

CO UPLE S

-50 C to 1760 C

50 C to 2000 C

CO UPLE B

200 C to 1820 C

50 C to 2000 C

CO UPLE E

-250

C to 1000

C

20 C to 1000 C

CO UPLE N

-250

C to 1300 C

20 C to 1000 C

COUPLE W5

0 C to 2320 C

 

50 C to 2000 C

 

 

 

 

 

The accuracy of the thermocouples is indicated in the appendix

Compensation of the cold junction of the thermocouples J,K,T,S,N,E,W5: 1.25 C

16.1.5.Sampling

 

Resolution:

14 bits

Max. sampling period:

 

Memory mode:

1 s (i.e. 1 MHz)

Direct mode:

200 s (i.e. 5 kHz)

Max. sampling duration:

10 min

16.1.6. Bandwidth

 

Bandwith at -3 dB:

 

Range

Bandwidth

> 1 V

100 kHz

> 50 mV

50 kHz

20 mV

30 kHz

10 mV

30 kHz

5 mV

20 kHz

Internal analogical filters:

10 kHz ,1 kHz, 100 Hz, 10Hz

slope:

20 dB/decade

Software filters:

1 Hz, 0,1 Hz, 0,01 Hz, 0,001 Hz

Frequency record:

 

Sensitivity

100 mVrms min.

Minimum duty factor

10 %.

Frequency

between 10 Hz and 100 kHz.

Accuracy

0.2 % of the full scale

Page 2

Page 121
Image 121
Escort 3008B, 3016B, 3004B user manual Technical specifications Temperature recording, Sampling, Bandwidth