![](/images/new-backgrounds/1005959/5959123x1.webp)
8Direct mode
Stop: condition for stopping the plotting
-Manual: with the Record
key
-Trigger: with an association of the analogical or logical channels see Chapter 6 Trigger
-Wait for : after a delay or at a precise date and time
-Paper length: after writing a programmed length of paper
After end of plot: action after the end of the plotting; only possible if the start and stop triggers are not Manual
-Stop: no action
-Rearm: waiting for the start trigger
-Change Setup.: loading a configuration and waiting for the start trigger
Save Data: simultaneous saving of the measurements in internal memory or on a file
-No save: no simultaneous saving
-Memory: recording in internal memory
-Acquisition period: sampling speed of the channels
-Blocks: dividing the internal memory into blocks
-File: recording on a file
-File name: place and name of the saving file
-Acquisition period: sampling speed of the channels
See chapter File mode to know the limitations of the file recording
Trigger: condition for stopping the simultaneous saving in internal memory only
-Direct: the same as for paper printing
-Manual: with the Record
key
-Trigger: with an association of the analogical or logical channels see Chapter Trigger
-
Following: actions after the end of the simultaneous acquisition into the memory
-Save on file: saving the acquisition into a file
-Plot: writing the simultaneous acquisition into the memory
-Rearm.: reloading the writing; waiting for the start condition again
If a speed before start has been programmed, the paper will be fed at this rate while waiting the trigger.
The user can urge the trigger at any time by pressing the "Start/Stop" key again while waiting for the start
While plotting, pressing this key will stop the plotting
In case of power failure while plotting, the recorder will start again in the same status (waiting or writing)
Page 4