4 Use
4.5. Channels set up key
Configuration of the channels. After pressing this key, select a module, then select a channel to access its parameters.
4.5.1. Analogical channels
Name: give a name to the channel (max. 26 characters)
Type: choice of the type of measure made on this input
-voltage, intensity, frequency or thermocouple
-direct or RMS
-value of the shunt in intensity measurement
-choice of the type in thermocouple, compensation, unit
Filter: positioning a filter at the inlet
-10 kHz, 1 kHz, 100 Hz, 10 Hz for analogical filters
-1 Hz, 10 s, 100 s or 1000 s for digital filters
Function: makes it possible to assign a mathematical calculation function to the considered channel
-Without: no function.
-Change unit: modifies the unit of the measures on the channel; you can then program a couple of dots X1, Y1 and X2, Y2 to achieve a scaling
-Calculation: available mathematical functions, associated parameters and unit