Procedure 1
1.Remove all bar codes from the laser beam path.
2.Place the supplied calibration symbol in the laser beam in the same position (distance and angle) as the labels to be verified will be scanned.
3.Press and hold the RESET button on the SV unit until the Calibration LED begins to flash.
4.Release the RESET button immediately after the Calibration LED begins flashing.
If calibration is successful, the laser beam will go off and the Calibration LED will go off.
5.Remove the calibration symbol from the beam path, and press the RESET button until the beam turns on. The SV unit is now ready to operate.
If calibration is unsuccessful, the Calibration LED will be either on steadily or flashing. In this case, repeat the calibration procedure.
Procedure 2
1.Place the supplied calibration symbol in the laser beam in the same position (distance and angle) as the labels to be verified will be scanned.
2.Send “~SC” into the SV unit serial port. This will cause the Calibration LED to flash.
3.If calibration is successful, the laser beam will go off and the Calibration LED will go off. Remove the calibration symbol from the beam path and then Press the RESET button until the beam turns on. The SV unit is now ready to operate.
4.If calibration is unsuccessful, the Calibration LED will be either on steadily or flashing. In this case, repeat the calibration procedure.
Once calibration is successful,
Quick Check® SV Series User’s Guide | 5 - 9 |