1.8 Power Line Connection Data

Power lead-in must be routed over a 30 mA fault current interrupter that is provided by the customer. The room installation must comply with VDE 0107.

In all other countries outside the Federal Republic Germany, the legally specified country regulations take precedence and must be maintained.

Prerequisite:

The unit is designed for single-phase DC current with a fixed installation and for a fixed connection using an all-pole separator from the Network (IEC 601, Chapter 57.1).

During installation, it must be possible to adapt the power line voltage and frequency to correspond to the order.

Power connection:

1 N 115

/200/208/ 230/240 V AC

Frequency:

50 / 60 Hz

Nominal current (fuse):

13/7.5/7.2/6.5/6.25 A

Nominal line power:

1500 VA

Heat dissipation:

240

W

1.9 Required Special Tools

Torque wrench

50 Nm (5 mkp)

Masonry drill bit

12 mm dia.

Sista sanitary caulking

F 101

Special grease - Tunap Tungrease BS

Pa. Nr.: 9026 0001

1.10 Required Test Equipment

PC or laptop with color display (black/white is also sufficient, but operation is made more difficult);

min. 486 processor; 16 MB RAM; with a 3 ½" diskette drive and hard drive.

Windows 3.x ; Win 95 operating system (Windows NT can cause difficulties with the Com interface). Serial connection cable (9-pole, Sub-D connector to 9-pole Sub-D socket, 1-1 connection), length > 1m.

Spirit level

Tape measure

Multimeter test instrument

1.11 Conditions for Transport and Storage

Ambient temperature range

-25 C to 70 C

Relative humidity in the range

5% to 95%

Air pressure in the range

700 hPa to 1100 hPa

1.12 Operating Conditions

Ambient temperature range

10 C to 40 C

Relative humidity in the range

20% to 80%

Air pressure in the range

700 hPa to 1100 hPa

06/00

- 11 of 90 -

0116 7201

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf -Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru