6. Board Replacement

All boards are connected to each other by connectors and are replaceable. After they have been replaced correctly, they should perform a self test after power is switched on and log themselves into the CPU.

The CPU board and breaker board are accessible after opening the URO cover panel. They are seated on several round snap-on mounts, each of which musts be pressed through. The footswitch board is accessible only after opening the metal housing.

The control unit with spiral must be completely replaced.

The display with connection cable must be completely replaced.

Do not disconnect plug-in connections that are under voltage!

7. Version Update

Only the software of the CPU board can be updated to the latest version. The software in the other board cannot be updated.

The update is performed by replacing the EPROM on the processor module.

EPROM's are CMOS components. Please observe the guidelines for EMI protection.