Tabletop transverse movement:

-Even movement?

-Soft startup and soft outward movement?

-Automatic stop in the centered position?

-Automatic stop in the limit position?

-Movement initiated from footswitch?

-Movement initiated from the manual control unit?

Movable Tube Unit Support Arm:

-Easy separation in the exposure position?

-Easy separation in the park position?

-Even movement?

-Positive engagement into the limit positions?

Table extension:

-Does the table extension engage in place in the opening at the foot end?

-Is the table extension blocked from sliding out?

-Does the controller when the table extension is controlled?

Footrest:

- Does the footrest securely and easily engage the footrest?

Micturation seat:

-Does the micturation seat engage in place in the opening?

-Is the micturation seat secured against falling out?

-Does the controller sense the micturation seat?

Accessories:

- Are all configured accessories functional?

General:

-Is the centering of the tube unit to the adjustable Bucky okay?

-Is the centering of the tube unit to the I.I. okay?

-Is radiation blocked when the tube unit support arm is in the park position?

06/00

- 71 of 90 -

0116 7201

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf -Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru