Fig.61

Fig.62

3.8 Potentiometer 1 – Table Tilt Movement Path

Take off the potentiometer belt (Fig. 61/Pos. 2). Move the vertical carriage into the top position and tilt the tabletop so that the potentiometer moves into the highest possible position, e.g. -20° position. Turn potentiometer =AU -R1 (Pos. 1) until it is half a turn in front of the mechanical end position; while doing this, take note of the direction of the turn. Place the potentiometer belt back on without turning the potentiometer and secure it in position.

3.9Potentiometer 2 – Table Tilt Movement Path

Take off the potentiometer belt (Fig. 62/Pos. 2). Move the vertical carriage into the top position and tilt the tabletop so that potentiometer =AU -R3 (Pos. 1) moves into the highest possible position, e.g. +88° position. Turn the potentiometer until it is half a turn in front of the mechanical end position; while doing this, take note of the direction of the turn. Place the potentiometer belt back on without turning the potentiometer and secure it in position.

06/00

- 53 of 90 -

0116 7201

Rev. 00 © 2000

Hans Pausch Röntgengerätebau Graf -Zeppelin-Str. 1

D-91056 Erlangen ALL RIGHTS RESERVED Ru