3.23 Operation in the Service Mode

a. Connection of a Laptop

To establish a connection with the URO controller, a PC/laptop is required.

Equipment:

PC or laptop with color display (black/white is also sufficient, but makes operation more difficult)

486 processor, 16 MB RAM

Windows 3.x Win 95 operating system (Windows NT can cause difficulties with the Com interface)

Hard drive

Serial connection cable (9-pole, Sub-D connector to a 9-pole, Sub-D socket, connected 1 - 1), length > 1m.

Current version of the "hp-Service" software

b. Procedure

Switch off the URO

Connect the PC and the URO CPU board using the connection cable

Switch on the URO

After approx. 1 s, the service LED will begin to blink

Switch on the PC

Start the PC software

Set the PC software to monitor. The terminal screen will appear

Set the service switch on the CPU board to the up "Service" position

After approx. 1 s, the service LED will switch to continuously on

The start message of the service menu will appear on the screen.

U

U

RRRR

OOO

333

000

000

000

U U

R R

O O

 

3 0 00 0 00 0 00

U

U

RRRR

O O

33

0 0 0 0 0 0 0 0 0

U U

R R

O O

 

3 00 0 00 0 00 0

 

UUU

R R

OOO

333

000

000

000

 

 

 

Service mode

 

 

 

 

 

(1)

Input test

 

 

 

 

 

(2)

Analog test

 

 

 

 

 

(3)

Handcontrol and footswitch test

 

 

(4)

Axis calibration

 

 

 

 

(5)

Versioncheck

 

 

Select testmode

06/00

- 59 of 90 -

0116 7201

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf -Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

Page 59
Image 59
Hansol Multitech Uromat 3000 installation instructions Operation in the Service Mode, Equipment