Hansol Multitech Uromat 3000 installation instructions Installing the Grid Replacing the Grid

Models: Uromat 3000

1 90
Download 90 pages 15.61 Kb
Page 46
Image 46

2.16 Installing the Grid – Replacing the Grid

Lift up the tabletop (CFC panel) (Fig. 27/Pos. 1), it will be held by the gas spring.

Move the tabletop towards the back and remove the front Nirosta panel. Move the tabletop towards the front and remove the rear Nirosta panel. Remove the left (Fig. 27/Pos. 3) and right (Pos. 4) cover panels. Remove the screws for the spiral cable guide panel and remove the cable guide panel. Move the tabletop longitudinally, remove the 6 mounting screws in the polycarbonate panel (Pos. 2) and remove the polycarbonate panel towards the front.

Switch the unit to the service mode, the green LED on the controller will go on and stay on. Press the Store and Reset buttons on the controller simultaneously, the three memory LED’s on the controller will blink green and the LED on the controller will blink very rapidly.

The cassette Bucky can now be moved slowly forward and backward using the Recall and Reset buttons.

Move the Bucky into the exposure position using the Recall button. Remove the mounting brackets (Fig. 28/ Pos. 2). Insert the grid with the focus mark facing up and secure it in place with the mounting brackets. Reinstall the polycarbonate panel (Fig. 27Pos. 2).

2.17 Installing the Tilt Angle Display- Error Display

Place the unit base cover panel (Fig. 29/Pos. 1) on the unit base. Guide the cable (Pos. 2) into the conduit (Pos. 3), install the tilt angle display (Pos. 4) and plug in the connector. Secure the plug-in connector with cable ties.

Fig. 29

Fig. 30

0116 7201

-46 of 90 -

06/00

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf -Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

Page 46
Image 46
Hansol Multitech Uromat 3000 Installing the Grid Replacing the Grid, Installing the Tilt Angle Display- Error Display