Fig.63

Fig.64

3.10Tabletop Longitudinal Safety Switch

Lift up the tabletop (it will be held in place by the gas spring) and remove the cover panel (Fig. 63/Pos. 1). Set the software end positions to 240+/- 5 mm in each direction. Adjust the limit switch (Fig. 64/Pos. 2) so that it switches off approx. 3 - 5

mmbehind each end position. To do this, either the limit switch or the switch strike place can be adjusted. Check the function over the entire movement path and reinstall the cover panel.

3.11 Tabletop Transverse Safety Switch

not configured

Fig.65

0116 7201

Rev. 00 © 2000 Hans Pausch Röntgengerätebau

Fig.66

 

-54 of 90 -

 

06/00

Graf -Zeppelin-Str. 1

D-91056 Erlangen

ALL RIGHTS RESERVED Ru

Page 54
Image 54
Hansol Multitech Uromat 3000 Tabletop Longitudinal Safety Switch, Tabletop Transverse Safety Switch Not configured