4.2 Function Check

Switch on the system. Initiate movement through the entire movement path.

Vertical drive:

-Even movement?

-Automatic stop in the bottom and top positions?

-Soft start and smooth movement out?

-Movement initiated from footswitch?

-Movement initiated from manual control unit?

-Any uninitiated lowering with load applied?

Tilt drive:

-Even movement?

-Soft start and smooth movement out?

-Automatic stop in vertical limit position (88°) ?

-Automatic stop in horizontal limit position (0°) ?

-Automatic stop in Trendelenburg position (-20°) ?

-Movement initiated from footswitch?

-Movement initiated from manual control unit?

-Any uninitiated lowering with load applied?

System longitudinal movement:

-Even movement?

-Soft start and smooth movement out?

-Automatic stop in head-end limit position?

-Automatic stop in foot-end limit position?

-Movement initiated from footswitch?

-Movement initiated from manual control unit?

-Any movement with opened Bucky flap?

Tabletop longitudinal movement:

-Even movement?

-Soft bootup and soft outward movement?

-Automatic stop in the middle position?

-Automatic stop in the head-end limit position?

-Automatic stop in the foot-end limit position?

-Movement initiated from footswitch?

-Movement initiated from manual control unit?

0116 7201

-70 of 90 -

06/00

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf -Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

Page 70
Image 70
Hansol Multitech Uromat 3000 installation instructions Function Check