4.2 Function Check
Switch on the system. Initiate movement through the entire movement path.
Vertical drive:
-Even movement?
-Automatic stop in the bottom and top positions?
-Soft start and smooth movement out?
-Movement initiated from footswitch?
-Movement initiated from manual control unit?
-Any uninitiated lowering with load applied?
Tilt drive:
-Even movement?
-Soft start and smooth movement out?
-Automatic stop in vertical limit position (88°) ?
-Automatic stop in horizontal limit position (0°) ?
-Automatic stop in Trendelenburg position
-Movement initiated from footswitch?
-Movement initiated from manual control unit?
-Any uninitiated lowering with load applied?
System longitudinal movement:
-Even movement?
-Soft start and smooth movement out?
-Automatic stop in
-Automatic stop in
-Movement initiated from footswitch?
-Movement initiated from manual control unit?
-Any movement with opened Bucky flap?
Tabletop longitudinal movement:
-Even movement?
-Soft bootup and soft outward movement?
-Automatic stop in the middle position?
-Automatic stop in the
-Automatic stop in the
-Movement initiated from footswitch?
-Movement initiated from manual control unit?