3.12Bucky Movement Path Safety Switch

Move the Bucky to the foot end to the mechanical stop, move the system carriage into the exposure position. Adjust the S12 switch (Fig. 66/Pos. 1) so that it positively switches.

3.13 I.I. Carriage Movement Path Safety Switch

Remove the cover panel (Fig. 65/Pos. 1). Move the system carriage to the head end until there is 3 mm distance and adjust the switch strike plate (Fig. 67/ Pos. 1) so that it just actuates the S6 switch. Move the system carriage to the foot end until there is 3 mm distance and adjust the switch strike plate (Fig. 67/Pos. 2) so that it just actuates the S7 switch.

Fig.67

Fig.68

3.14 Table Up Movement Path Switch Strike Plate

Raise the table to max. 1225+/-10mm position.

When the software limit is set, the top switch strike plate (Fig. 68 / Pos. 1) can be adjusted so that the limit switch (Pos. 3) actuates approx. 2 - 5 mm behind the maximum height.

3.15 Table Down Movement Path Switch Strike Plate

Lower the table to the min. 680+/-10mm position.

When the software limit is set, the bottom switch strike plate (Fig. 68/ Pos. 2) can be adjusted so that the limit switch actuates approx. 2 - 5 mm behind the minimum height.