-Check automatic stop in foot-end limit position

-Clean the guide rails.

-Check for play in the bearings. If necessary, readjust.

Tube unit support arm, exposure position, park position:

-Press the buttons on the control handle and move through the entire movement range.

-Check for even movement during the unit movements.

-Check for engagement into the end positions.

-Clean the guide rails.

-Check for play in the bearings. If necessary, readjust them.

-Check for fluoroscopy and exposure blocking when the tube unit support arm is not in the exposure position.

Component Panel:

-Switch power off to the system.

-Check all cables laid for good condition and secure connection. If necessary, replace any damaged cables.

-Retighten any cable connections.

-Retighten the mounting screws of all components.

-Check P/C boards for secure seating.

-Check strain-reliefs and ground connections for secure connection.

Electronics:

- Electronic adjustment parameters per section.3.28

Accessories:

-Check all accessories present at the customer’s location.

-Hook in the table extension, tilt the table and check the factory set safety distance of 12 cm; if necessary, adjust it.

-Check engagement of the leg support on the profile rail.

-Check for secure mounting of the head cushion mount.

-Check the patient table pad for good condition.

-Check the flush bowl for damage and check the function.

General:

-Reinstall the unit base cover panel.

-Reinstall the unit base cover panels.

-Reinstall the base cover panels.

-Reinstall the system carriage cover panels.

-Clean the unit.

-Clean surfaces with one of the following tested disinfecting agents:

Tego 103, Kosolin, Misty Multi-Purpose Disinfectant Cleaner, Misty Multi-Purpose Disinfectant Cleaner II, Misty Disinfectant and Deodorant RTU, Precise Hospital Foam Cleaner Disinfectant.

-Switch power back on.

-Carry out a function check according to 4.2.

06/00

- 69 of 90 -

0116 7201

Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf -Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

Page 69
Image 69
Hansol Multitech Uromat 3000 installation instructions