5.Troubleshooting
Error | Cause | Corrective Measure | ||
Number | ||||
|
| |||
F | 1 | Frequency converter 0 for tilt/vertical, | Check connection cables, settings | |
|
| system failed. | (address, Baud rate). | |
F | 2 | Frequency converter 1 for tilt/vertical, table | Check connection cables, settings | |
|
| longitudinal movement, cassette failed. | (address, Baud rate). | |
F | 3 | Frequency converters 0+1 failed. | Check connection cables, settings | |
|
|
| (address, Baud rate). | |
|
|
| ||
|
|
| ||
|
|
| ||
|
|
| ||
F 10 | Footswitch communication problem. | Check fuse, cabling. | ||
F 11 | Manual control unit communication | Check fuse, cabling. | ||
|
| problem. |
| |
|
|
| ||
F 13 | Incorrect footswitch software version. | Replace footswitch. | ||
F 14 | Incorrect manual control unit software | Replace manual control unit. | ||
|
| version. |
| |
|
|
| ||
|
|
| ||
|
|
| ||
F 20 | Analog voltage failed. | Check fuse. | ||
F 21 | M1 tilt/vertical potentiometer error. | Check potentiometer cables and | ||
|
|
| potentiometer. | |
F 22 | System potentiometer error. | Check potentiometer cables and | ||
|
|
| potentiometer. | |
F 23 | M3 tilt/vertical potentiometer error. | Check potentiometer cables and | ||
|
|
| potentiometer. | |
F 24 | Table longitudinal potentiometer error. | Check potentiometer cables and | ||
|
|
| potentiometer. | |
F 25 | Cassette potentiometer error. | Check potentiometer cables and | ||
|
|
| potentiometer. | |
F26 | Table transverse potentiometer error / | Check motor potentiometer cables | ||
|
| motor defective. | and potentiometer. | |
|
|
| ||
F 28 | Collimator vertical potentiometer error. | Check potentiometer cables and | ||
|
|
| potentiometer. | |
|
|
| ||
F 30 | Button fuse malfunction. | Check buttons, cables. | ||
F 31 | Footswitch button sticking. | Check buttons. | ||
F 32 | Manual control unit button sticking. | Check buttons. | ||
F 33 | Exposure switch active during system start. | Check switches. | ||
F 34 | Fluoroscopy switch active during system | Check switches. | ||
|
| start. |
| |
F 35 | Exposure centering switch defective. | Check switches, cabling. | ||
|
|
| ||
|
|
| ||
|
|
| ||
0116 7201 | 06/00 |
Rev. 00 © 2000 Hans Pausch Röntgengerätebau Graf