Fig.55

Fig.56

3.4Movement Path Potentiometer, Tabletop longitudinal

Lift up the tabletop (Fig. 55/Pos. 2) (it will be held in place by the gas spring) and remove the cover panel (Fig. 55/Pos. 1). Move potentiometer =AU -R4 (Fig. 56/Pos.

1)away from the pinion rack, move the tabletop into the centered position. Turn the potentiometer to the middle position, mesh it back into the pinion rack and align the pinion on the rack. Check the function over the entire movement path and reinstall the cover panel.

Fig.57

06/00

Rev. 00 © 2000

Fig.58

 

- 51 of 90 -

0116 7201

Hans Pausch Röntgengerätebau Graf -Zeppelin-Str. 1 D-91056 Erlangen ALL RIGHTS RESERVED Ru

Page 51
Image 51
Hansol Multitech Uromat 3000 installation instructions Movement Path Potentiometer, Tabletop longitudinal