Adjusting the Angle of Vision

The viewing angle can be adjusted from 5° to 32° by tilting the binocular eyepiece.

8.4 Eyepieces

Note:

The eyepiece’s aperture protector must be re- moved during microscopy while wearing eye- glasses, or at least folded back.

Eyeglasses with multifocal lenses (bifocals and smooth view glasses) must be removed while operating the macroscope.

Correction for Vision Problems

With your right eye, look through the right eyepiece and bring the specimen into sharp focus.

If both oculars have an adjustable eyelens, the eyelens of the right ocular can be brought to the position for normal vision, which is in- dicated by an encircling line.

When using a graticule in the right eyepiece, the eyelens must first be focused on the grat- icule.

Then, view the same specimen using the left eye and rotate the eyelens of the left eye- piece until the object appears in sharp focus. While doing so, do not use the focus dial.

8.5Brightness Setting Cold Light Illumination

The control dials (20.8 and 20.12) adjust the first and (if applicable) the second cold light source. The function is dependent on the cabling on the rear side of the stand page 19.

We recommend using the same cold light source for both sides based on the illumination method being used. The second cold light source should then be used for another illumina- tion method.

Illumination method page 26ff.

Note:

The LED (20.9) or (20.11) is green if the corre- sponding lamp is operated in standard mode. If the LED is red, the bulb life of the lamp is reduced.

UV Illumination

The UV lamps are equipped with a high-pres- sure, 180 W quartz radiator and UV filter disks for a wavelength of 366 nm. They are rated for a direct 230 V/50 Hz power supply.

For further details see the special instruc- tions.

8.6 Focusing

You can focus using the handwheels on the left and right side (20.1 and 20.7).

If the button (20.10) is pressed, i.e. if the syn- chronous adjustment of the stages is activated, focusing is also synchronous.