VXI 3002 manual Internal ΦM flatness test, Internal AM depth and distortion tests at −4 dBm

Models: 3002

1 136
Download 136 pages 1.74 Kb
Page 120
Image 120

Table 5-29 Internal ΦM flatness test

Modulation

Response

Result (dB)

Response

frequency (kHz)

level min.

 

level max.

 

(dB)

 

(dB)

 

 

 

 

 

 

 

 

0.1

3

_______

+3

0.3

3

_______

+3

1

reference

3

3

_______

+3

10

3

_______

+3

 

 

 

 

 

 

 

 

Table 5-30 Internal AM depth and distortion tests at 4 dBm

 

AM depth 30%

AM depth 80%

Distortion

 

 

 

 

 

 

 

 

 

Carrier

min.

Result

max.

min.

Result

max.

Result at

Result at

frequency

(%)

(%)

(%)

(%)

(%)

(%)

30% depth

80% depth

(MHz)

 

 

 

 

 

 

(<1.5%)

(<2.5%)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.5

28.5

_____

31.5

76

_____

84

_______

_______

6

28.5

_____

31.5

76

_____

84

_______

_______

9

28.5

_____

31.5

76

_____

84

_______

_______

11

28.5

_____

31.5

76

_____

84

_______

_______

20

28.5

_____

31.5

76

_____

84

_______

_______

50

28.5

_____

31.5

76

_____

84

_______

_______

100

28.5

_____

31.5

76

_____

84

_______

_______

200

28.5

_____

31.5

76

_____

84

_______

_______

500

28.5

_____

31.5

76

_____

84

_______

_______

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 5-31 Internal AM depth and distortion tests at +7 dBm

 

AM depth 30%

AM depth 80%

Distortion

 

 

 

 

 

 

 

 

 

Carrier

min.

Result

max.

min.

Result

max.

Result at

Result at

frequency

(%)

(%)

(%)

(%)

(%)

(%)

30% depth

80% depth

(MHz)

 

 

 

 

 

 

(<1.5%)

(<2.5%)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1.5

28.5

_____

31.5

76

_____

84

_______

_______

6

28.5

_____

31.5

76

_____

84

_______

_______

9

28.5

_____

31.5

76

_____

84

_______

_______

11

28.5

_____

31.5

76

_____

84

_______

_______

20

28.5

_____

31.5

76

_____

84

_______

_______

50

28.5

_____

31.5

76

_____

84

_______

_______

100

28.5

_____

31.5

76

_____

84

_______

_______

200

28.5

_____

31.5

76

_____

84

_______

_______

500

28.5

_____

31.5

76

_____

84

_______

_______

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5-45

Page 120
Image 120
VXI 3002 manual Internal ΦM flatness test, Internal AM depth and distortion tests at −4 dBm