(3)Measure the distortion on the audio analyzer checking that the result is within the specification shown in Table 5-44.

(4)Measure the absolute level on the audio analyzer (in dBm) and record this level as a reference.

(5)Set the UUT mod source to each of the frequencies shown in Table 5-44. Subtract the level measured on the audio analyzer at each frequency from that recorded in (4) checking that the results are within specification.

External frequency standard input

Specification

External input

Front-panel BNC connector accepts an input of 1 MHz or 10 MHz at 220 mV RMS to

 

1.8 V RMS into 1 k.

Test equipment

 

Description

Minimum specification

Example

Signal generator

220 mV to 1.8 V RMS, 1 MHz to 10 MHz

IFR 2041 or 2030

Test procedure

 

Signal generator

UUT

 

FREQ STD

RF OUTPUT

 

I/O

 

 

C2570

Fig. 5-15 External standard test set-up.

(1)Connect the test equipment as shown in Fig. 5-15.

(2)On the UUT set:

Freq Std

External 1 Direct

(3)Set the signal generator to RF level 220 mV EMF, carrier frequency 1 MHz.

(4)Using Table 5-45, check that no external standard error messages are displayed on the UUT.

(5)Set the signal generator to 1.8 V EMF and repeat (4).

(6)On the UUT set

Freq Std

External 10 Indirect

(7)Set the signal generator to carrier frequency 10 MHz and repeat (4).

(8)Set the signal generator to 220 mV and repeat (4).

5-22

Page 97
Image 97