VXI 3002 manual Attenuator accuracy, Measuring

Models: 3002

1 136
Download 136 pages 1.74 Kb
Page 80
Image 80

Attenuator accuracy

Test procedure

 

 

Signal generator

UUT

 

RF OUTPUT

RF

 

 

OUT

RF

LO

 

 

Down converter

 

Frequency offset

Sensor

 

mode connection

 

 

 

Measuring

 

 

receiver

 

 

C2559

Fig. 5-2 RF output test set-up

(1)Connect the test equipment as shown in Fig. 5-2.

(2)On the UUT set:

Carr freq

2.6 MHz

RF Level

−4 dBm

Set ∆

11 ENTER

(3)Tune the receiver to 2.6 MHz and record the output level measured in Table 5-9 checking that the result is within specification.

(4)Set the UUT RF level to −4.1 dBm. Measure the received level and record the result in Table 5-9, checking that the result is within specification.

(5)Decrement the UUT, using the ￿ icon, in 11 dB steps down to an RF level of −103.1 dBm measuring the received level at each step shown in Table 5-9 checking that the results are within specification.

(6)Set the UUT to carrier frequency 480.1 MHz and repeat (2) to (5) using Table 5-10.

(7)Set the UUT to carrier frequency 1199 MHz and repeat (2) to (5) using Table 5-11.

The down converter will automatically be enabled when testing frequencies above 1300 MHz.

(8)Set the local oscillator to +8 dBm at a carrier frequency of 62 MHz less than the test frequency (i.e. 1813.1 MHz).

(9)On the receiver, enter the local oscillator frequency followed by the test frequency.

(10)Set the UUT to carrier frequency 1875.1 MHz and repeat (2) to (5) using Table 5-12.

(11)Set the UUT to carrier frequency 2399 MHz and repeat (2) to (5) using Table 5-13.

5-5