VXI 3002 manual Attenuator hold, Attenuator lock ranges, Atten?

Models: 3002

1 136
Download 136 pages 1.74 Kb
Page 57
Image 57

Miscellaneous commands

Attenuator hold

The ATTEN:LOCK command allows you to reduce the RF level by at least another 10 dB without the step attenuator operating.

ATTEN

[not used alone]

:LOCK

Lock the Attenuators

:UNLOCK

Unlock the Attenuators

Data type :

None

 

Allowed suffices :

None

 

Default suffix :

None

 

Example:

ATTEN:LOCK

ATTEN?

Prepares message containing information on whether the Attenuators are

 

locked or unlocked in the following format:

 

:ATTEN:<status>

 

where

<status> is a program mnemonic indicating whether the

 

 

attenuators are locked or unlocked.

Example: :ATTEN:LOCK

Table 3-2 below applies to software versions 1.03 onwards. Maximum level with attenuator lock on will reduce if AM or pulse modulation is applied. Any user programmed limits for RF level or offsets will also affect the attenuator lock ranges.

Table 3-2 Attenuator lock ranges

 

RF level

Atten. lock

Atten. lock

setting (dBm)

(min. dBm)

(max. dBm)

 

 

 

 

 

from

 

to

 

 

 

 

 

 

 

+26.0

 

+18.0

+7

+25

+17.9

 

+7.1

4

+24

+7.0

 

+7.0

4

+13

+6.9

 

-4.0

15

+13

4.1

 

15.0

26

+2

15.1

 

26.0

37

9

26.1

 

37.0

48

20

37.1

 

48.0

59

31

48.1

 

59.0

70

42

59.1

 

70.0

81

53

70.1

 

81.0

82

54

81.1

 

92.0

103

75

92.1

 

103.0

114

86

103.1

 

114.0

125

97

114.1

 

125.0

136

108

125.1

 

136.0

137

119

136.1

 

137.0

137

130

 

 

 

 

 

 

 

 

 

 

3-27

Page 57
Image 57
VXI 3002 manual Attenuator hold, Attenuator lock ranges, RF level Atten. lock Setting dBm Min. dBm Max. dBm From, Atten?