RF level frequency response

Test procedure

UUT

 

6960B

 

RF power meter

 

6932

SENSOR

RF OUT

INPUT

 

 

Power

 

 

sensor

 

 

 

C2558

Fig. 5-1 RF output test set-up

(1)Perform AUTO ZERO and AUTO CAL on the power meter.

(2)Connect the test equipment as shown in Fig. 5-1.

(3)On the UUT set:

Carr Freq

30 kHz

RF Level

0 dBm

(4)Record the output level measured by the power meter against each of the carrier frequencies shown in Table 5-1 checking that the results are within specification.

(5)Set the UUT RF level to +7 dBm and repeat (4) using Table 5-2.

(6)Set the UUT RF level to +25 dBm and repeat (4) using Table 5-3, decreasing the RF level to +19 dBm when testing at carrier frequencies above 1.2 GHz.

ALC linearity

Test procedure

(1)Perform AUTO ZERO and AUTO CAL on the power meter.

(2)Connect the test equipment as shown in Fig. 5-1.

(3)On the UUT set:

Carr Freq

2.5 MHz

RF Level

−4 dBm

(4)Record the output level measured by the power meter against each of the steps shown in Table 5-4 checking that the results are within specification.

(5)Set the UUT carrier frequency to 950 MHz and repeat (4) using Table 5-5.

(6)Set the UUT carrier frequency to 1200 MHz and repeat (4) using Table 5-6.

(7)Set the UUT carrier frequency to 1900 MHz and repeat (4) using Table 5-7.

(8)Set the UUT carrier frequency to 2400 MHz and repeat (4) using Table 5-8.

5-4

Page 79
Image 79
VXI 3002 manual RF level frequency response, ALC linearity, Test procedure