TEST PROCEDURES
Each test procedure relies on the UUT being set to its
Store/Recall Address 999
Recall
At the end of this chapter are a set of results tables which give all the test points for each of the tests. These tables should be photocopied and used to record the results of all the measurements taken.
RF output
Specification
Range | −137 dBm to +25 dBm (+19 dBm above 1.2 GHz). |
|
|
| ||||
|
| When AM is selected, the maximum RF output level decreases linearly with increasing | ||||||
|
| AM depths to +19 dBm (+13 dBm above 1.2 GHz) at 99% depth. | ||||||
Accuracy |
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
| |
|
|
|
| Accuracy over temperature range 17°C to 27°C |
| |||
|
|
|
|
| 9 kHz to 1.2 GHz | 1.2 GHz to 2.4 GHz |
| |
|
|
|
| >−127 dBm | ±1.0 dB | ±2.0 dB |
| |
|
|
|
|
|
|
|
| |
|
|
|
| Temperature coeff. over temperature range 0°C to 55°C |
| |||
|
|
|
|
| 9 kHz to 1.2 GHz | 1.2 GHz to 2.4 GHz |
| |
|
|
|
|
| <±0.02 dB/°C | <±0.04 dB/°C |
| |
VSWR |
|
|
|
|
|
|
| |
For output levels less than −5 dBm output VSWR is less than 1.3:1 for carrier | ||||||||
|
| frequencies up to 1.2 GHz and less than 1.5:1 for carrier frequencies up to 2.4 GHz. | ||||||
Output impedance | 50 Ω SMA female connector to MIL 390123D. |
|
|
| ||||
Test equipment |
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
| Description |
|
| Minimum specification |
| Example | ||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| Power meter |
| ±0.1 dB from 9 kHz to 2.4 GHz |
| IFR 6960B and | |||
|
|
|
|
|
|
| 6932 |
|
| Measuring |
| 0 dBm to −127 dBm; 2.5 MHz to 2.4 GHz |
| Agilent 8902A | |||
| receiver |
|
|
|
|
| with 11722A | |
|
|
|
|
|
|
| sensor and | |
|
|
|
|
|
|
| 11793A down | |
|
|
|
|
|
|
| converter | |
| Signal generator | +8 dBm from 32.5 MHz to 2.43 GHz |
| IFR 2041 | ||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|