Carrier error
Test procedure
(1)Connect the test equipment as shown in Fig.
(2)On the UUT set:
Carr Freq | 1200 MHz |
RF Level | 0 dBm |
(3)On the modulation meter select CARRIER ERROR. The FREQUENCY display will read 0.00 kHz.
(4)On the UUT set:
Mod Mode | FM External |
FM1 Level | 100 kHz |
Source On |
|
Mod On |
|
DC coupling |
|
Null DCFM |
|
(5)On the modulation meter measure the carrier frequency error displayed in the FREQUENCY window checking that the result is within the specification shown in Table
External FM frequency response (ALC off, DC coupled)
Test procedure
| UUT | 2305 |
| Modulation meter | |
Function generator |
| RF |
| OUT | |
| EXT MOD | RF |
OUTPUT | INPUT | INPUT |
|
| LF |
Temporary |
| OUTPUT |
connection | 50 Ω |
|
DVM | load |
|
| Audio analyzer | |
|
| |
|
| INPUT |
|
| C2565 |
Fig. 5-8 External modulation and modulation distortion test set-up
30 Hz to 100 kHz
(1)Connect the test equipment as shown in Fig.
(2)On the UUT set:
Carr Freq | 15 MHz |
RF Level | 0 dBm |
Mod Mode | FM External |
FM1 Level | 50 kHz |
Source On |
|
Mod On |
|
DC coupling |
|
(3)Set the function generator to give 1V RMS, 1 kHz sine wave.
(4)On the modulation meter select CAL, FM, 10 Hz 300 kHz filter.