Pulse modulation RF level frequency response
Fig. A-1 Pulse modulation test set-up
Test procedure
(1)Perform AUTO ZERO and AUTO CAL on the power meter.
(2)Connect the test equipment as shown in Fig.
(3)On the UUT set:
Carr Freq | 100 kHz |
RF Level | −7 dBm |
Pulse ON |
|
(4)Set the function generator to provide +5 V DC. The RF output will now be enabled.
(5)Record the output level measured by the power meter against each of the carrier frequencies shown in Table
(6)Set the UUT RF level to +0 dBm and repeat (5) using Table
(7)Set the UUT RF level to +16 dBm and repeat (5) using Table
Pulse modulation on/off ratio
| UUT |
| RF |
Function generator | OUT |
| |
OUTPUT |
|
| PULSE |
| INPUT |
| Spectrum analyzer |
| RF INPUT |
| C5358 |