Pulse modulation RF level frequency response

Fig. A-1 Pulse modulation test set-up

Test procedure

(1)Perform AUTO ZERO and AUTO CAL on the power meter.

(2)Connect the test equipment as shown in Fig. A-1.

(3)On the UUT set:

Carr Freq

100 kHz

RF Level

−7 dBm

Pulse ON

 

(4)Set the function generator to provide +5 V DC. The RF output will now be enabled.

(5)Record the output level measured by the power meter against each of the carrier frequencies shown in Table A-1, checking that the results are within specification.

(6)Set the UUT RF level to +0 dBm and repeat (5) using Table A-2.

(7)Set the UUT RF level to +16 dBm and repeat (5) using Table A-3.

Pulse modulation on/off ratio

 

UUT

 

RF

Function generator

OUT

 

OUTPUT

 

 

PULSE

 

INPUT

 

Spectrum analyzer

 

RF INPUT

 

C5358

Fig. A-2 Pulse modulation on/off ratio test set-up

Annex-A-2

Page 128
Image 128
VXI 3002 manual Test procedure, Spectrum analyzer