Annex A

OPTION 11 FAST PULSE MODULATION

General description

Option 11 adds the ability for the instrument to internally generate a fast pulse modulated waveform from logic levels applied to the PULSE INPUT socket. The pulse modulator is suitable for generating fast pulses with high isolation for applications in radar and EMI. It may be used in conjunction with other forms of modulation to form composite signals. Familiarity with normal operation of the signal generator is assumed.

Operation

Fast pulse modulation is automatically applied when Option 11 is fitted.

Acceptance testing

The following acceptance tests supersede those given under Pulse modulation in Chapter 5.

Pulse modulation

Specification

Pulse modulation

 

Carrier frequency range

100 kHz to 2.4 GHz, usable to 9 kHz.

RF output range

As standard instrument, with the exception that maximum output level

 

is reduced by 3 dB when pulse modulation is enabled.

RF level accuracy

Additional ±0.01 dB/°C temperature coefficient when pulse modulation

 

enabled.

 

Adds ±0.25 dB for carrier frequencies below 10 MHz.

On/off ratio

>80 dBc below 1.2 GHz;

 

>70 dBc up to 2.05 GHz (typically >80 dB);

 

>65 dBc up to 2.4 GHz (typically >70 dB at 2.4 GHz).

Rise and fall time

<20 ns, typically 15 ns.

Maximum repetition frequency

10 MHz.

Control

50 input impedance. A logic ‘1’ (2 to 5 V) turns the carrier on, a

 

logic ‘0’ (0 to 0.8 V) turns the carrier off. Maximum input is ±10 V.

Test equipment

 

Alternative equipment may be used providing it complies with the stated minimum performance.

Description

Minimum specification

Example

 

 

 

Power meter

±0.1 dB from 9 kHz to 2.4 GHz

IFR 6960B and

 

 

6912

Spectrum

Frequency coverage 100 kHz to 2.4 GHz

AnritsuMS2602A

analyzer

 

 

50 load

1 W, 50 nominal impedance, DC to 2.4 GHz

Lucas Weinschel

(termination)

 

M1404N

Oscilloscope

100 MHz bandwidth

Tektronix

 

 

TDS 220

Function

DC to 10 kHz square wave

HP 3325B

generator

 

 

RF detector

100 kHz to 2 GHz

HP 8471D

 

 

 

Annex-A-1